CarboVis 70x IQ TS Technical data
8 - 1
ba75950e02 11/2012
8 Technical data
8.1 Measurement characteristics
Measuring principle Spectrophotometric absorption measurement; integrated micropro-
cessor electronics, shielded 2-wire connection for
power and data transmission.
Wavelength range 200 - 720 nm
Measuring gap (optical
layer thickness)
Carbon measurement,
1 mm gap width
CarboVis 701 IQ TS 1 mm
CarboVis 705 IQ TS 5 mm
Measuring
location
Measuring mode Measuring range Resolution
Inlet COD spectr., total
0 ... 20000 mg/L 1 mg/L
COD spectr., dissolv
0 ... 12500 mg/L 1 mg/L
TOC spectr., correl.
0 ... 20000 mg/L 1 mg/L
DOC spectr., correl.
0 ... 12500 mg/L 1 mg/L
BOD spectr., correl.
0 ... 8000 mg/L 1 mg/L
SAC 254, total
0 ... 5000 1/m 1 1/m
SAC 254, dissolved
0 ... 3000 1/m 1 1/m
UVT 254
0.0 ... 100.0 % 0.1 %
Activation COD spectr., dissolv
0 ... 12500 mg/L 1 mg/L
DOC spectr., correl.
0 ... 12500 mg/L 1 mg/L
SAC 254, total
0 ... 5000 1/m 1 1/m
SAC 254, dissolved
0 ... 3000 1/m 1 1/m
UVT 254
0.0 ... 100.0 % 0.1 %