Vue de la page 22
CarboVis 70x IQ TS Commissioning
3 - 13
ba75950e02 11/2012
Fig. 3-7 140 - Settings of sensors and diff. sensors (virtual sensor)
3.4.5 Settings of the secondary sensor (TSS measurement)
Carrying out settings Using <S>, switch from the measured value display to the main menu
of the settings. Then navigate to the setting menu (setting table) of the
sensor. The procedure is described in detail in your IQ S
ENSOR NET
system operating manual.
Default values are marked in bold.
Save and quit -
The system confirms the saving of the
settings and the display switches to the
next higher level.
Quit -
The display switches to the next higher
level without saving the new settings.
Menu item Settings Explanations
Menu item Settings Explanations
Measuring mode TSS
Display of the measuring mode
(TSS = total suspended solids measure-
ment)
Measuring range
-
Display of the measuring range
(see chapter 8 T
ECHNICAL DATA).
Cal - # raw value 1
Value pairs of the user calibration for TSS measurement (see section
4.3.2).
Note:
Leave those values unchanged during the initial commission-
ing.
Cal - ref. value 1
Cal - # raw value 2
Cal - ref. value 2