CarboVis 70x IQ TS Commissioning
3 - 5
ba75950e02 11/2012
3.3.3 Connecting the sensor to the IQ SENSOR NET
Connection cable A sensor connection cable of the SACIQ or SACIQ SW type is required
to connect the sensor. The cable is available in different lengths.
Compared to the standard model SACIQ, the SACIQ SW sensor
connection cable is optimized regarding its corrosion resistance in
seawater and brackish water and adapted for use in conjunction with
the CarboVis 70x IQ TS . Information on this and other IQ S
ENSOR NET
accessories is given in the WTW catalog and on the Internet.
How to connect the SACIQ (SW) sensor connection cable to the termi-
nal strip of an MIQ module is described in chapter 3 I
NSTALLATION of the
IQ S
ENSOR NET system operating manual.
Are the plug connections
dry?
Before connecting the sensor and sensor connection cable, please
make sure the plug connections are dry. If moisture gets into the plug
connections, first dry the plug connections (dab them dry or blow them
dry using compressed air).
Note
Do not suspend the sensor on the sensor connection cable but use the
suitable mounting equipment. Information on this and other
IQ S
ENSOR NET accessories is given in the WTW catalog and on the In-
ternet.