CarboVis 70x IQ TS Technical data
8 - 3
ba75950e02 11/2012
Carbon measurement,
5 mm gap width
TSS measurement,
1 mm gap width
TSS measurement,
5 mm gap width
The quoted measuring ranges are nominal measuring ranges that are
theoretically possible. In practice, real measuring ranges exist that are
given by the limits of photometric determination. The limits are signifi-
cantly influenced by the light scattering due to solids and the absorption
of accompanying substances (sample matrix). Therefore, no fixed
measuring range can be determined beforehand.
8.2 Application conditions
Suitability and areas of
application
Municipal wastewater with a low proportion of industrial wastewater
Measuring medium:
Measuring
location
Measuring mode Measuring range Resolution
Outlet COD spectr., total
0.0 ... 800.0 mg/L 0.1 mg/L
COD spectr., dissolv
0.0 ... 800.0 mg/L 0.1 mg/L
TOC spectr., correl.
0.0 ... 500.0 mg/L 0.1 mg/L
DOC spectr., correl.
0.0 ... 500.0 mg/L 0.1 mg/L
BOD spectr., correl.
0.0 ... 500.0 mg/L 0.1 mg/L
SAC 254, total
0.0 ... 600.0 1/m 0.1 1/m
SAC 254, dissolved
0.0 ... 600.0 1/m 0.1 1/m
UVT 254
0.0 ... 100.0 % 0.1 %
Measuring
location
Measuring
mode
Measuring range Resolution
Inlet TSS
0.00 ... 15.00 g/L 0.01 g/L
Outlet TSS
0 ... 4500 mg/L 1 mg/L
Measuring
location
Measuring
mode
Measuring range Resolution
Outlet TSS
0.0 ... 900.0 mg/L 0.1 mg/L
pH value 4 - 12
Flow rate < 3 m/s
Temperature 0 °C ... + 45 °C
(32 °F ... + 113 °F)