OPERATIONS MANUALba75950e02 11/2012CarboVis 701 IQ TSCarboVis 705 IQ TSOPTICAL SENSOR FOR CARBON SUM PARAMETERS AND TOTAL SUSPENDED SOLIDS (TSS)
Safety CarboVis 70x IQ TS2 - 2ba75950y01 11/20122.2 Safe operation2.2.1 Authorized useThe authorized use of the CarboVis 70x IQ TS consists of its use
CarboVis 70x IQ TS Commissioning3 - 1ba75950e02 11/20123 Commissioning 3.1 IQ SENSOR NET system requirementsSoftware statuses of thecontroller and ter
Commissioning CarboVis 70x IQ TS3 - 2ba75950e02 11/20123.3 Installation3.3.1 Mounting the sensorNoteThe sensor is an optical precision instrument that
CarboVis 70x IQ TS Commissioning3 - 3ba75950e02 11/2012Vertical installation For vertical installation, mount the sensor in a position that allows int
Commissioning CarboVis 70x IQ TS3 - 4ba75950e02 11/20123.3.2 Mounting the shock protectorsThe shock protectors consist of two rings, a cap and four ca
CarboVis 70x IQ TS Commissioning3 - 5ba75950e02 11/20123.3.3 Connecting the sensor to the IQ SENSOR NETConnection cable A sensor connection cable of t
Commissioning CarboVis 70x IQ TS3 - 6ba75950e02 11/2012Connecting the sensor tothe sensor connectioncableFig. 3-5 Connecting the sensor1. Take the pro
CarboVis 70x IQ TS Commissioning3 - 7ba75950e02 11/20123.4 Initial commissioning3.4.1 General informationCAUTIONNever look into the measuring window o
Commissioning CarboVis 70x IQ TS3 - 8ba75950e02 11/20123.4.2 Sensor structureThe physical sensor is the sensor with the instrument or model desig-nati
CarboVis 70x IQ TS Commissioning3 - 9ba75950e02 11/20123.4.3 Settings for the main sensorMeasuring cycle A measuring cycle consists of the cleaning pr
2 ba75950e02 11/2012CarboVis 70x IQ TSFor the most recent version of the manual, please visit www.ysi.com.Contact YSI1725 Brannum LaneYellow Springs,
Commissioning CarboVis 70x IQ TS3 - 10ba75950e02 11/2012Menu item Settings ExplanationsMeasuring modeCOD spectr., totalCOD spectr., dissolvTOC spec
CarboVis 70x IQ TS Commissioning3 - 11ba75950e02 11/2012Response time t90 10 ... 60 minResponse time of the signal smoothing.Depending on the sample m
Commissioning CarboVis 70x IQ TS3 - 12ba75950e02 11/20123.4.4 Settings for virtual sensorsCarrying out settings Using <S>, switch from the measu
CarboVis 70x IQ TS Commissioning3 - 13ba75950e02 11/2012Fig. 3-7 140 - Settings of sensors and diff. sensors (virtual sensor)3.4.5 Settings of the sec
Commissioning CarboVis 70x IQ TS3 - 14ba75950e02 11/2012Fig. 3-8 140 - Settings of sensors and diff. sensors (Secondary sensor)Main sensorInformation
CarboVis 70x IQ TS Measuring / Operation4 - 1ba75950e02 11/20124 Measuring / Operation4.1 Determination of measured valuesFig. 4-1 Determination of me
Measuring / Operation CarboVis 70x IQ TS4 - 2ba75950e02 11/20124.2 Measuring operationCAUTIONNever look into the measuring window or put objects into
CarboVis 70x IQ TS Measuring / Operation4 - 3ba75950e02 11/20124.3 Calibration4.3.1 OverviewIn the delivery condition, the sensor is precalibrated and
Measuring / Operation CarboVis 70x IQ TS4 - 4ba75950e02 11/2012Sensor check (H2O dest.)The characteristics of a sensor and thus the measured values pr
CarboVis 70x IQ TS Measuring / Operation4 - 5ba75950e02 11/20124.3.2 User calibrationCalibration points andvalue pairsThe adjustment is carried out at
CarboVis 70x IQ TS Contents0 - 1ba75950e02 11/2012CarboVis 70x IQ TS - Contents1 Overview . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
Measuring / Operation CarboVis 70x IQ TS4 - 6ba75950e02 11/2012Performing a usercalibrationProceed as follows if you want to perform a user calibratio
CarboVis 70x IQ TS Measuring / Operation4 - 7ba75950e02 11/2012Default values are marked in bold.Calibration evaluation The entered calibration data a
Measuring / Operation CarboVis 70x IQ TS4 - 8ba75950e02 11/20124.3.3 Sensor check/Zero adjustmentNoteThe sensor check or Zero adjustment must be carri
CarboVis 70x IQ TS Measuring / Operation4 - 9ba75950e02 11/2012FlowchartFig. 4-3Overview of Sensortest/Zero adjustment and user calibration.After clea
Measuring / Operation CarboVis 70x IQ TS4 - 10ba75950e02 11/2012Preparing the sensorcheck orZero adjustmentPrepare the sensor check or Zero adjustment
CarboVis 70x IQ TS Measuring / Operation4 - 11ba75950e02 11/2012Fig. 4-5 Putting the sensor sleeve on the sensorCarry out the intended sensor check or
Measuring / Operation CarboVis 70x IQ TS4 - 12ba75950e02 11/2012Up to this point, you can break off the calibration procedure at any time with the <
CarboVis 70x IQ TS Measuring / Operation4 - 13ba75950e02 11/2012If no valid zero adjustment is possible under the current conditions, you can restore
Measuring / Operation CarboVis 70x IQ TS4 - 14ba75950e02 11/2012
CarboVis 70x IQ TS Maintenance and cleaning5 - 1ba75950e02 11/20125 Maintenance and cleaning5.1 MaintenanceThe UV-VIS CarboVis 70x IQ TS sensor operat
CarboVis 70x IQ TS0 - 2ba75950y01 11/20125 Maintenance and cleaning . . . . . . . . . . . . . . . . . . . . . . .5-15.1 Maintenance . . . . . . . .
Maintenance and cleaning CarboVis 70x IQ TS5 - 2ba75950e02 11/20125.2.2 General steps to be takenDepending on the application site and the level of co
CarboVis 70x IQ TS Maintenance and cleaning5 - 3ba75950e02 11/20125.2.3 Basic cleaningSteps of the basiccleaningAfter completing the cleaning procedur
Maintenance and cleaning CarboVis 70x IQ TS5 - 4ba75950e02 11/20125.2.4 Cleaning the measuring gapTo remove lime or grease deposits, use the flocked c
CarboVis 70x IQ TS Spare parts, maintenance equipment, accessories6 - 1ba75950e02 11/20126 Spare parts, maintenance equipment, accessoriesGeneral acce
Spare parts, maintenance equipment, accessories CarboVis 70x IQ TS6 - 2ba75950e02 11/2012Mounting accessoriesInformation on other IQ SENSOR NET access
CarboVis 70x IQ TS What to do if...7 - 1ba75950e02 11/20127 What to do if..."----" display(invalid measured value) Implausible measuredvalu
What to do if... CarboVis 70x IQ TS7 - 2ba75950e02 11/2012Drifting measured valuesDisplay of OFL Measured valuesfluctuating heavily Cause Remedy– Cl
CarboVis 70x IQ TS What to do if...7 - 3ba75950e02 11/2012Measured values too lowor too high Display of ERRORCause Remedy– Gas bubbles in front of th
What to do if... CarboVis 70x IQ TS7 - 4ba75950e02 11/2012
CarboVis 70x IQ TS Technical data8 - 1ba75950e02 11/20128 Technical data8.1 Measurement characteristicsMeasuring principle Spectrophotometric absorpti
CarboVis 70x IQ TS Overview1 - 1ba75950e02 11/20121Overview1.1 How to use this component operating manualStructure of theIQ SENSOR NET operatingmanual
Technical data CarboVis 70x IQ TS8 - 2ba75950e02 11/2012Outlet COD spectr., total0 ... 4000 mg/L 1 mg/LCOD spectr., dissolv0 ... 4000 mg/L 1 mg/LTOC s
CarboVis 70x IQ TS Technical data8 - 3ba75950e02 11/2012Carbon measurement,5 mm gap widthTSS measurement,1 mm gap widthTSS measurement,5 mm gap widthT
Technical data CarboVis 70x IQ TS8 - 4ba75950e02 11/2012Pressure resistance Sensor with connected SACIQ (SW) sensor connection cable:The sensor meets
CarboVis 70x IQ TS Technical data8 - 5ba75950e02 11/20128.3 General dataDimensions(in mm) WeightConnection technique Connection via SACIQ (SW) sensor
Technical data CarboVis 70x IQ TS8 - 6ba75950e02 11/2012Instrument safetyTest certificates 8.4 Electrical data Applicable norms – EN 61010-1– UL 6101
CarboVis 70x IQ TS Indexes9 - 1ba75950e02 11/20129 Indexes9.1 Explanation of the messagesThis chapter contains a list of all the message codes and rel
Indexes CarboVis 70x IQ TS9 - 2ba75950e02 11/20129.1.1 Error messagesMessage code Message textEA2541Sensor temperature too high!* Check process and ap
CarboVis 70x IQ TS Indexes9 - 3ba75950e02 11/2012ES5541Failure optical measurement: hardware* Contact serviceES6541Failure optical measurement: softwa
Indexes CarboVis 70x IQ TS9 - 4ba75950e02 11/20129.1.2 Info messagesMessage code Message textIAC541Zero adjustment was successfully carried out* Carry
CarboVis 70x IQ TS Indexes9 - 5ba75950e02 11/20129.2 Status infoThe status info is a coded piece of information on the current status of a sensor. Eac
Overview CarboVis 70x IQ TS1 - 2ba75950e02 11/20121.2 Fields of applicationThe range of application comprises control, feedback control and mon-itorin
Indexes CarboVis 70x IQ TS9 - 6ba75950e02 11/2012
CarboVis 70x IQ TS Appendix: Glossary10 - 1ba75950e02 11/201210 Appendix: GlossaryResolution Smallest difference between two measured values that can
Appendix: Glossary CarboVis 70x IQ TS10 - 2ba75950e02 11/2012Test sample Designation of the test sample ready to be measured. Measuring gap The measur
CarboVis 70x IQ TS Contact Information11 - 1ba75950e02 11/201211 Contact Information11.1 Ordering & Technical SupportWhen placing an order please
Contact Information CarboVis 70x IQ TS11 - 2ba75950e02 11/2012
1) The tissue in plants that brings water upward from the roots;2) a leading global water technology company.We're 12,500 people unified in a com
CarboVis 70x IQ TS Overview1 - 3ba75950e02 11/20121.4 Structure of the UV-VIS sensorA UV-VIS sensor contains a light source (1), a measuring gap (3) t
Overview CarboVis 70x IQ TS1 - 4ba75950e02 11/2012
CarboVis 70x IQ TS Safety2 - 1ba75950y01 11/20122Safety2.1 Safety information2.1.1 Safety information in the operating manualThis operating manual pro