Page 1 - CarboVis 705 IQ TS

OPERATIONS MANUALba75950e02 11/2012CarboVis 701 IQ TSCarboVis 705 IQ TSOPTICAL SENSOR FOR CARBON SUM PARAMETERS AND TOTAL SUSPENDED SOLIDS (TSS)

Page 2 - © 2012 Xylem Inc

Safety CarboVis 70x IQ TS2 - 2ba75950y01 11/20122.2 Safe operation2.2.1 Authorized useThe authorized use of the CarboVis 70x IQ TS consists of its use

Page 3 - CarboVis 70x IQ TS - Contents

CarboVis 70x IQ TS Commissioning3 - 1ba75950e02 11/20123 Commissioning 3.1 IQ SENSOR NET system requirementsSoftware statuses of thecontroller and ter

Page 4 - CarboVis 70x IQ TS

Commissioning CarboVis 70x IQ TS3 - 2ba75950e02 11/20123.3 Installation3.3.1 Mounting the sensorNoteThe sensor is an optical precision instrument that

Page 5 - 1Overview

CarboVis 70x IQ TS Commissioning3 - 3ba75950e02 11/2012Vertical installation For vertical installation, mount the sensor in a position that allows int

Page 6 - 1.2 Fields of application

Commissioning CarboVis 70x IQ TS3 - 4ba75950e02 11/20123.3.2 Mounting the shock protectorsThe shock protectors consist of two rings, a cap and four ca

Page 7 - CarboVis 70x IQ TS Overview

CarboVis 70x IQ TS Commissioning3 - 5ba75950e02 11/20123.3.3 Connecting the sensor to the IQ SENSOR NETConnection cable A sensor connection cable of t

Page 8 - Overview CarboVis 70x IQ TS

Commissioning CarboVis 70x IQ TS3 - 6ba75950e02 11/2012Connecting the sensor tothe sensor connectioncableFig. 3-5 Connecting the sensor1. Take the pro

Page 9 - 2.1 Safety information

CarboVis 70x IQ TS Commissioning3 - 7ba75950e02 11/20123.4 Initial commissioning3.4.1 General informationCAUTIONNever look into the measuring window o

Page 10 - 2.2 Safe operation

Commissioning CarboVis 70x IQ TS3 - 8ba75950e02 11/20123.4.2 Sensor structureThe physical sensor is the sensor with the instrument or model desig-nati

Page 11 - 3 Commissioning

CarboVis 70x IQ TS Commissioning3 - 9ba75950e02 11/20123.4.3 Settings for the main sensorMeasuring cycle A measuring cycle consists of the cleaning pr

Page 12 - 3.3 Installation

2 ba75950e02 11/2012CarboVis 70x IQ TSFor the most recent version of the manual, please visit www.ysi.com.Contact YSI1725 Brannum LaneYellow Springs,

Page 13 - Flow direction

Commissioning CarboVis 70x IQ TS3 - 10ba75950e02 11/2012Menu item Settings ExplanationsMeasuring modeCOD spectr., totalCOD spectr., dissolvTOC spec

Page 14

CarboVis 70x IQ TS Commissioning3 - 11ba75950e02 11/2012Response time t90 10 ... 60 minResponse time of the signal smoothing.Depending on the sample m

Page 15 - NSTALLATION of the

Commissioning CarboVis 70x IQ TS3 - 12ba75950e02 11/20123.4.4 Settings for virtual sensorsCarrying out settings Using <S>, switch from the measu

Page 16 - SACIQ (SW)

CarboVis 70x IQ TS Commissioning3 - 13ba75950e02 11/2012Fig. 3-7 140 - Settings of sensors and diff. sensors (virtual sensor)3.4.5 Settings of the sec

Page 17 - 3.4 Initial commissioning

Commissioning CarboVis 70x IQ TS3 - 14ba75950e02 11/2012Fig. 3-8 140 - Settings of sensors and diff. sensors (Secondary sensor)Main sensorInformation

Page 18 - Signal smoothing

CarboVis 70x IQ TS Measuring / Operation4 - 1ba75950e02 11/20124 Measuring / Operation4.1 Determination of measured valuesFig. 4-1 Determination of me

Page 19 - Number of C-sensors

Measuring / Operation CarboVis 70x IQ TS4 - 2ba75950e02 11/20124.2 Measuring operationCAUTIONNever look into the measuring window or put objects into

Page 20 - ECHNICAL DATA)

CarboVis 70x IQ TS Measuring / Operation4 - 3ba75950e02 11/20124.3 Calibration4.3.1 OverviewIn the delivery condition, the sensor is precalibrated and

Page 21

Measuring / Operation CarboVis 70x IQ TS4 - 4ba75950e02 11/2012Sensor check (H2O dest.)The characteristics of a sensor and thus the measured values pr

Page 22

CarboVis 70x IQ TS Measuring / Operation4 - 5ba75950e02 11/20124.3.2 User calibrationCalibration points andvalue pairsThe adjustment is carried out at

Page 23

CarboVis 70x IQ TS Contents0 - 1ba75950e02 11/2012CarboVis 70x IQ TS - Contents1 Overview . . . . . . . . . . . . . . . . . . . . . . . . . . . . .

Page 24

Measuring / Operation CarboVis 70x IQ TS4 - 6ba75950e02 11/2012Performing a usercalibrationProceed as follows if you want to perform a user calibratio

Page 25 - 4 Measuring / Operation

CarboVis 70x IQ TS Measuring / Operation4 - 7ba75950e02 11/2012Default values are marked in bold.Calibration evaluation The entered calibration data a

Page 26 - 4.2 Measuring operation

Measuring / Operation CarboVis 70x IQ TS4 - 8ba75950e02 11/20124.3.3 Sensor check/Zero adjustmentNoteThe sensor check or Zero adjustment must be carri

Page 27 - 4.3 Calibration

CarboVis 70x IQ TS Measuring / Operation4 - 9ba75950e02 11/2012FlowchartFig. 4-3Overview of Sensortest/Zero adjustment and user calibration.After clea

Page 28

Measuring / Operation CarboVis 70x IQ TS4 - 10ba75950e02 11/2012Preparing the sensorcheck orZero adjustmentPrepare the sensor check or Zero adjustment

Page 29

CarboVis 70x IQ TS Measuring / Operation4 - 11ba75950e02 11/2012Fig. 4-5 Putting the sensor sleeve on the sensorCarry out the intended sensor check or

Page 30

Measuring / Operation CarboVis 70x IQ TS4 - 12ba75950e02 11/2012Up to this point, you can break off the calibration procedure at any time with the <

Page 31

CarboVis 70x IQ TS Measuring / Operation4 - 13ba75950e02 11/2012If no valid zero adjustment is possible under the current conditions, you can restore

Page 32

Measuring / Operation CarboVis 70x IQ TS4 - 14ba75950e02 11/2012

Page 33 - Flowchart

CarboVis 70x IQ TS Maintenance and cleaning5 - 1ba75950e02 11/20125 Maintenance and cleaning5.1 MaintenanceThe UV-VIS CarboVis 70x IQ TS sensor operat

Page 34 - by 90°

CarboVis 70x IQ TS0 - 2ba75950y01 11/20125 Maintenance and cleaning . . . . . . . . . . . . . . . . . . . . . . .5-15.1 Maintenance . . . . . . . .

Page 35 - Measuring

Maintenance and cleaning CarboVis 70x IQ TS5 - 2ba75950e02 11/20125.2.2 General steps to be takenDepending on the application site and the level of co

Page 36

CarboVis 70x IQ TS Maintenance and cleaning5 - 3ba75950e02 11/20125.2.3 Basic cleaningSteps of the basiccleaningAfter completing the cleaning procedur

Page 37

Maintenance and cleaning CarboVis 70x IQ TS5 - 4ba75950e02 11/20125.2.4 Cleaning the measuring gapTo remove lime or grease deposits, use the flocked c

Page 38

CarboVis 70x IQ TS Spare parts, maintenance equipment, accessories6 - 1ba75950e02 11/20126 Spare parts, maintenance equipment, accessoriesGeneral acce

Page 39 - 5 Maintenance and cleaning

Spare parts, maintenance equipment, accessories CarboVis 70x IQ TS6 - 2ba75950e02 11/2012Mounting accessoriesInformation on other IQ SENSOR NET access

Page 40 - Zero adjustment

CarboVis 70x IQ TS What to do if...7 - 1ba75950e02 11/20127 What to do if..."----" display(invalid measured value) Implausible measuredvalu

Page 41

What to do if... CarboVis 70x IQ TS7 - 2ba75950e02 11/2012Drifting measured valuesDisplay of OFL Measured valuesfluctuating heavily Cause Remedy– Cl

Page 42

CarboVis 70x IQ TS What to do if...7 - 3ba75950e02 11/2012Measured values too lowor too high Display of ERRORCause Remedy– Gas bubbles in front of th

Page 43 - ENSOR NET system 182

What to do if... CarboVis 70x IQ TS7 - 4ba75950e02 11/2012

Page 44

CarboVis 70x IQ TS Technical data8 - 1ba75950e02 11/20128 Technical data8.1 Measurement characteristicsMeasuring principle Spectrophotometric absorpti

Page 45 - 7 What to do if

CarboVis 70x IQ TS Overview1 - 1ba75950e02 11/20121Overview1.1 How to use this component operating manualStructure of theIQ SENSOR NET operatingmanual

Page 46

Technical data CarboVis 70x IQ TS8 - 2ba75950e02 11/2012Outlet COD spectr., total0 ... 4000 mg/L 1 mg/LCOD spectr., dissolv0 ... 4000 mg/L 1 mg/LTOC s

Page 47 - ENSOR NET cable to

CarboVis 70x IQ TS Technical data8 - 3ba75950e02 11/2012Carbon measurement,5 mm gap widthTSS measurement,1 mm gap widthTSS measurement,5 mm gap widthT

Page 48

Technical data CarboVis 70x IQ TS8 - 4ba75950e02 11/2012Pressure resistance Sensor with connected SACIQ (SW) sensor connection cable:The sensor meets

Page 49 - 8 Technical data

CarboVis 70x IQ TS Technical data8 - 5ba75950e02 11/20128.3 General dataDimensions(in mm) WeightConnection technique Connection via SACIQ (SW) sensor

Page 50

Technical data CarboVis 70x IQ TS8 - 6ba75950e02 11/2012Instrument safetyTest certificates 8.4 Electrical data Applicable norms – EN 61010-1– UL 6101

Page 51 - 8.2 Application conditions

CarboVis 70x IQ TS Indexes9 - 1ba75950e02 11/20129 Indexes9.1 Explanation of the messagesThis chapter contains a list of all the message codes and rel

Page 52

Indexes CarboVis 70x IQ TS9 - 2ba75950e02 11/20129.1.1 Error messagesMessage code Message textEA2541Sensor temperature too high!* Check process and ap

Page 53 - 8.3 General data

CarboVis 70x IQ TS Indexes9 - 3ba75950e02 11/2012ES5541Failure optical measurement: hardware* Contact serviceES6541Failure optical measurement: softwa

Page 54 - 8.4 Electrical data

Indexes CarboVis 70x IQ TS9 - 4ba75950e02 11/20129.1.2 Info messagesMessage code Message textIAC541Zero adjustment was successfully carried out* Carry

Page 55 - 9 Indexes

CarboVis 70x IQ TS Indexes9 - 5ba75950e02 11/20129.2 Status infoThe status info is a coded piece of information on the current status of a sensor. Eac

Page 56 - Indexes CarboVis 70x IQ TS

Overview CarboVis 70x IQ TS1 - 2ba75950e02 11/20121.2 Fields of applicationThe range of application comprises control, feedback control and mon-itorin

Page 57 - Message code Message text

Indexes CarboVis 70x IQ TS9 - 6ba75950e02 11/2012

Page 58

CarboVis 70x IQ TS Appendix: Glossary10 - 1ba75950e02 11/201210 Appendix: GlossaryResolution Smallest difference between two measured values that can

Page 59 - 9.2 Status info

Appendix: Glossary CarboVis 70x IQ TS10 - 2ba75950e02 11/2012Test sample Designation of the test sample ready to be measured. Measuring gap The measur

Page 60

CarboVis 70x IQ TS Contact Information11 - 1ba75950e02 11/201211 Contact Information11.1 Ordering & Technical SupportWhen placing an order please

Page 61 - 10 Appendix: Glossary

Contact Information CarboVis 70x IQ TS11 - 2ba75950e02 11/2012

Page 63 - 11 Contact Information

1) The tissue in plants that brings water upward from the roots;2) a leading global water technology company.We're 12,500 people unified in a com

Page 64

CarboVis 70x IQ TS Overview1 - 3ba75950e02 11/20121.4 Structure of the UV-VIS sensorA UV-VIS sensor contains a light source (1), a measuring gap (3) t

Page 65

Overview CarboVis 70x IQ TS1 - 4ba75950e02 11/2012

Page 66 - Xylem

CarboVis 70x IQ TS Safety2 - 1ba75950y01 11/20122Safety2.1 Safety information2.1.1 Safety information in the operating manualThis operating manual pro