Sharp JX-9400 Informations techniques Page 163

  • Télécharger
  • Ajouter à mon manuel
  • Imprimer
  • Page
    / 211
  • Table des matières
  • MARQUE LIVRES
  • Noté. / 5. Basé sur avis des utilisateurs
Vue de la page 162
Remember, however, that, in order to achieve good mixing, it is advanta-
geous to inject the tracer continuously whenever possible. A series of short
pulses spread evenly over time can simulate this continuous injection.
The air used to purge the pipes should ideally come from the measured zone
and be returned to the same zone. However, if the flow rate in the sampling
pipes is small, compared to the ventilation rate (which is the common case),
it is reasonable to purge the injection tubes with outside air and to exhaust
samples to the outside.
To ensure that the analysed air has a concentration averaged over the time
between two measurements; it is possible to continuously pump the air from
each zone into an inflatable bag that is then periodically emptied into the analyser.
Tracer gas analysers
Objective of the analysis
Measurement of the concentration of tracer in the air is the basic parameter
needed for the interpretation or control of the concentration itself. Several
analysing principles are available for such measurements. For any analyser,
the salient features to consider are the following:
.
Sensitivity it is desirable to minimize the quantity of tracer gas used, not
only from the point of view of cost, but also toxicity, fire or other hazards
that may be relevant. The more sensitive the analyser, the lower is the
required working concentration.
.
Selectivity the analyser shoul d not be sensitive to other gases usually
present in indoor air, for example, nitrogen, oxygen, water vapour, carbon
dioxide, argon and so on.
.
Speed the time needed for the analysis must be considered, especially
if several locations are to be sampled in sequence. The analysis time depends
on the type of instrument and varies from milliseconds for mass spectrometers
Tracer gas
Analysis
Purge
Purge
Tracer gas
Analysis
Pre-purge
Purge
injection
Figure 7.1 Two strategies for injection and sampling
Note: Left is one zone at a time; right is time shared.
142 Ventilation and Airflow in Buildings
Vue de la page 162
1 2 ... 158 159 160 161 162 163 164 165 166 167 168 ... 210 211

Commentaires sur ces manuels

Pas de commentaire