Sharp DM-4500 Instructions d'exploitation Page 25

  • Télécharger
  • Ajouter à mon manuel
  • Imprimer
  • Page
    / 164
  • Table des matières
  • DEPANNAGE
  • MARQUE LIVRES
  • Noté. / 5. Basé sur avis des utilisateurs
Vue de la page 24
25
6. Assembly
6.6 Light sources for the incident light axis
Caution!
Light sources pose a potential irradiation
risk (glare, UV-radiation, IR-radiation).
Therefore, lamps have to be operated in
closed housings.
Ensure that the lamp housing has been dis-
connected from the power supply. Unplug
the power plug and the power supply during
assembly.
During assembly work on xenon burners,
always wear the protective gloves and face
protection supplied (Fig. 15) (risk of explo-
sion).
Never touch the glass parts of the burner
with bare hands.
Never look directly into the beam path
(blinding hazard).
Lamp housing 106/106 z
This lamp housing is suitable for use with a 12 V
100 W halogen lamp or a variety of gas discharge
lamps.
Caution!
Make sure to follow the instructions and
safety notes of the lamp supplier.
Before changing lamps allow it to cool down
for at least 30 min.!
Fig. 16 106/106 z lamp housing (on the side, open)
1 Cover raised
2 Collector
3 12 V 100 W lamp or
gas discharge lamp in mount
4 Reflector (mirror)
5, 6, 7 Adjusting screw for x-y reflector
8 Fastening screw for the lamp mount
9 Socket for contact plug
1
2
3
898
4
5
6
7
Fig. 15
Protective gloves and mask
Vue de la page 24
1 2 ... 20 21 22 23 24 25 26 27 28 29 30 ... 163 164

Commentaires sur ces manuels

Pas de commentaire