Sharp DM-4500 Instructions d'exploitation Page 18

  • Télécharger
  • Ajouter à mon manuel
  • Imprimer
  • Page
    / 164
  • Table des matières
  • DEPANNAGE
  • MARQUE LIVRES
  • Noté. / 5. Basé sur avis des utilisateurs
Vue de la page 17
18
5. Unpacking the microscope
Installation location
Work with the microscope should be performed
in a dust-free room, which is free of oil vapors
and other chemical vapors, as well as extreme
humidity. At the workstation, large tempera-
ture fluctuations, direct sunlight and vibrations
should be avoided. These conditions can distort
measurements and micrographic images.
Allowable ambient conditions
Temperature 15-35°C
Relative humidity maximum 80% up to 30°C
Microscopes in warm and warm-damp climatic
zones require special care in order to prevent
the build up of fungus.
See the chapter on "Care of the microscope"
p. 77 for additional instructions.
Caution!
Electrical components must be assembled at
least 10 cm away from the wall and from
flammable substances.
Transport
For shipping or transporting the microscope
and its accessory components, the original
packaging should be used.
As a precaution to prevent damage from vibra-
tions, the following components should be dis-
assembled and packaged separately:
Unscrew the objectives.
Remove the condenser.
Remove the specimen stage.
Remove the lamp housings.
Disassemble the burner of 106 z lamp housing.
Remove all moving or loose parts.
Vue de la page 17
1 2 ... 13 14 15 16 17 18 19 20 21 22 23 ... 163 164

Commentaires sur ces manuels

Pas de commentaire