Sharp JX-9400 Informations techniques Page 164

  • Télécharger
  • Ajouter à mon manuel
  • Imprimer
  • Page
    / 211
  • Table des matières
  • MARQUE LIVRES
  • Noté. / 5. Basé sur avis des utilisateurs
Vue de la page 163
up to several minutes for gas chromatographs or multi-tracer infrared
analysers. Faster analysis will enable more frequent sampling of each zone
and hence provide more detailed data. Frequent sampling (for example,
every five to ten minutes) is essential for the constant concentration technique
to maintain accurate control of concentration.
.
Accuracy last but not least, the accuracy of the concentration measurement
directly influences the accuracy of the results.
There are several principles employed for analysis for tracer gas concentrations
that differ in the gases analysed, the ra nge of con centration detecte d, accuracy,
speed, ease of use and cost. These principles are discussed and specific
examples are given below.
Infrared absorption spectrometry
Any polyatomic gas molecule exhibits vibration modes, which are excited by
infrared radiation. The wavelengths of the infrared radiation correspondi ng
to molecular vibration frequencies
1
are absorbed in proportion to the number
of tracer gas molecules present in the infrared beam. This absorption is used
as a measure of the concentration of tracer gas molecules in the path between
an infrared source and detector. This technique is referred to as absorption
spectroscopy.
Infrared absorption spectrometers may be either dispersive or non-dispersive
types, and both are in common use. Dispersive spectrometers use a diffraction
grating that reflects the electromagnetic waves of a light beam into different
directions, each direction corresponding to a given wavelength. The instrument
is tuned by the operator to a narrow band of wavelengths specific to the gas of
interest and any convenient infrared detector measures the absorption. Some
modern instruments use tun eable laser diodes that emit the appropriate
wavelength.
In non-dispersive devices, all the infrared radiation present in the absorption
bands of the tracer is used. The infrared light beam is sent through both a refer-
ence channel containing nitrogen or pure air and an analysis channel that
contains a sample of room air. A chopper is used to alternately pass the radiation
from each channel to an analysis chamber containing a pure sample of the tracer
gas. This gas heats and cools in response to the modulated beam. The heated gas
expands through a measuring channel into an expansion chamber. The resulting
alternating flow through the measuring channel is measured by a highly sensitive
gas flow detector, which transmits an electric signal.
Characteristics of infrared absorption spectrometers:
.
Analysable tracers N
2
O, SF
6
, halon 1301, CO
2
. Other detectable gases,
such as H
2
O, benzene, alkenes and so on, are not suitable as tracers.
.
The sensitivity depends on the analysed gas. For example, full-scale devia-
tion is 200 ppm for N
2
O, or 20 ppm for SF
6
and CF
3
Br.
.
Interfering gases care must be taken to eliminate the effects of other gases
absorbing at similar frequencies (cross-sensitivity), particularly water
Common Methods and Techniques 143
Vue de la page 163
1 2 ... 159 160 161 162 163 164 165 166 167 168 169 ... 210 211

Commentaires sur ces manuels

Pas de commentaire