Sharp SYSTEM 8/10 Plus Manuel d'instructions Page 53

  • Télécharger
  • Ajouter à mon manuel
  • Imprimer
  • Page
    / 403
  • Table des matières
  • DEPANNAGE
  • MARQUE LIVRES
  • Noté. / 5. Basé sur avis des utilisateurs
Vue de la page 52
Installation
Setup
Thermo Fisher Scientific Model 5030i SHARP Instruction Manual 2-15
ambient T/RH sensor assembly. Tighten this adapter against the
stainless steel tubing using the two setscrews.
9. Sleeve the tripod over the tripod adapter to a proper height and tighten.
If using a tripod, using the adjustable legs of the tripod, plumb the inlet
tube assembly.
10. Attach the white ambient T/RH sensor radiation shield assembly onto
a portion of the exposed tripod adapter.
11. Attach and tighten the 5/8-inch to 3/8-inch reducing T-fitting to the
5/8-inch stainless steel tubing.
12. Place the PM
10
or PM
2.5
inlet into the 5/8-inch stainless T-fitting.
13. Feed one end of the black 25-foot T/RH cable trough he roof flange
assembly and connect to the bottom of the ambient T/RH assembly.
14. Connect the other end of the T/RH cable to the back of the instrument
(RH Temp).
15. Attach the heater power connector to HEATER 1.
16. Using the available 3/8-inch green vacuum tubing, connect the intake
of the pump to the vacuum fitting at the rear panel of the 5030i.
17. Plug the pump power chord into the AC PUMP receptacle at the rear
of the 5030i instrument.
Vue de la page 52
1 2 ... 48 49 50 51 52 53 54 55 56 57 58 ... 402 403

Commentaires sur ces manuels

Pas de commentaire