Operation
Alarms Menu
3-70 Model 5030i SHARP Instruction Manual Thermo Fisher Scientific
NEPHELOMETER TEMP:
ACTUAL MIN: 0.0
o
C
SET MIN TO: 5.0
o
C?
©ª INC/DEC
SAVE VALUE
RANGE AVG DIAGS ALARM
The Nephelometer RH screen displays the current nephelometer relative
humidity value and sets the minimum and maximum alarm limits.
Acceptable alarm limits range from 5 to 95%. If the nephelometer RH
reading goes beyond the minimum or maximum limit, an alarm is activated
and the alarm (bell) icon appears in the status bar on the Run screen and in
the Main Menu.
In the Main Menu, choose Alarms > Neph Alarms > Neph RH.
NEPHELOMETER RH:
ACTUAL 0.0 %
>MIN 5.0 %
MAX 95.0 %
RANGE AVG DIAGS ALARM
The Minimum Nephelometer RH alarm limit screen is used to change the
minimum nephelometer RH alarm limit. The minimum and maximum
nephelometer RH screens function the same way.
In the Main Menu, choose Alarms > Neph Alarms > Neph RH >
Min or Max.
NEPHELOMETER RH:
ACTUAL MIN: 5.0 %
SET MIN TO: 5.5 %?
©ª INC/DEC
SAVE VALUE
RANGE AVG DIAGS ALARM
The LED Current screen displays the current nephelometer source current
value and sets the minimum and maximum alarm limits. Acceptable alarm
limits range from 50 to 70 mA. If the LED current reading goes beyond
Nephelometer RH
Min and Max
Nephelometer RH Limits
LED Current