Sharp SYSTEM 8/10 Plus Manuel d'instructions Page 162

  • Télécharger
  • Ajouter à mon manuel
  • Imprimer
  • Page
    / 403
  • Table des matières
  • DEPANNAGE
  • MARQUE LIVRES
  • Noté. / 5. Basé sur avis des utilisateurs
Vue de la page 161
Operation
Service Menu
3-100 Model 5030i SHARP Instruction Manual Thermo Fisher Scientific
ANALOG OUTPUT CAL: ZERO
CONNECT METER TO OUTPUT!
SELECTED OUTPUT: V1
SET TO: 100
SAVE VALUE ©ª INC/DEC
SET OUTPUT TO: 0.0 V
RANGE AVG DIAGS ALARM
The Analog Output Calibrate Full-Scale screen allows the user to calibrate
the full-scale state of the selected analog output. The operator must connect
a volt meter to the output and adjust the output until the meter reads the
value shown in the “set output to:” field, in either V or mA, depending on
the selected output channel.
In the Main Menu, choose Service > Analog Out Cal > select
Channel > Calibrate Full Scale.
ANALOG OUTPUT CAL: SPAN
CONNECT METER TO OUTPUT!
SELECTED OUTPUT: V1
SET TO: 3397
SAVE VALUE ©ª INC/DEC
SET OUTPUT TO: 10 V
RANGE AVG DIAGS ALARM
The Analog Input Calibration menu is used to calibrate the 8 analog input
channels at both zero and full-scale. The analog input calibration menu is
visible only when the I/O expansion board is installed and when the
instrument is in service mode. For more information on the service mode,
see “Service Mode” earlier in this chapter.
Note This adjustment should only be performed by an instrument service
technician.
In the Main Menu, choose Service > Analog Input Calibration.
Analog Output Calibrate
Full-Scale
Analog Input Calibration
Vue de la page 161
1 2 ... 157 158 159 160 161 162 163 164 165 166 167 ... 402 403

Commentaires sur ces manuels

Pas de commentaire