Acquisition Interface
CS 9300 System User Guide (SM748)_Ed 01 3–15
“Select” 3D Acquisition Overview
The “Select” 3D Acquisition interface is the 3D interface for the CS 9300 that provides
you with a selected range of imaging acquisition functions.
Figure 3–14 “Select” 3D Acquisition interface
1 Information button:
About: Identifies the Software and the Firmware versions.
Reset of the Values: Resets to the manufacturing parameter settings.
Memorize settings: Memorizes the user preference settings for each patient type (kV,
mA and voxel size).
Save current mode as default: Sets the current selected program to be available by
default each time you access the Acquisition interface.
Activate the Phantom mode: Sets the unit parameters to perform an image quality
control test.
2 Preview Screen:
Before exposure: Displays an illustration of the selected FoV inside the anatomy.
During exposure: Displays the acquired image in real time.
3 Selected Parameter Display: Displays the current acquisition parameter settings.
4 System Status Screen: Displays various alert or warning messages originating from the unit.
5 Generator Cooling indicator: Indicates the automatic cooling time (mm:ss) required for the
generator to reach 0 for a new acquisition.