M4000 Insulation Analyzer User Guide
72A-1230 Rev. F 4-45
July 22, 2005
Select the Clipboard mode to perform the test and configure a test with
the potential of 500V. The test must be performed in Manual mode,
and any test circuit will do.
Raise the test voltage to 500V and record the DVM reading on the
Calibration Verification Certificate.
The setup for the test is shown in the following diagram: