M4000 Insulation Analyzer User Guide
72A-1230 Rev. F 4-45
July 22, 2005
• Select the Clipboard mode to perform the test and configure a test with
the potential of 500V. The test must be performed in Manual mode,
and any test circuit will do.
• Raise the test voltage to 500V and record the DVM reading on the
Calibration Verification Certificate.
The setup for the test is shown in the following diagram: